POC Glucose Meter Evaluated for Neonatal Intensive Care

By LabMedica International staff writers
Posted on 17 Oct 2013
The monitoring of blood glucose in neonatal intensive care unit (NICU) patients is important in maintaining normoglycemia and reducing the risk of hypoglycemia.

Point-of-care testing (POCT) glucose meters provide short turnaround times but some have been reported to be affected by hematocrit interference and other biochemical or biological substances in their accuracy and performance.

Scientists at the Medical University of Vienna (Austria) tested 159 heparinized whole blood samples from NICU patients obtained for blood gas analysis. Accuracy (bias) of a new meter and analytical interferences were evaluated by comparing the results of the meter with the results of the blood gas analyzer routinely used for glucose measurements in this NICU setting.

Routine glucose measurements were performed at each hospital site using blood gas analyzers ABL 700 and ABL 835 (Radiometer Medical ApS, Copenhagen, Denmark). The blood gas analyzers were considered the appropriate reference method for comparing the performance of StatStrip Glucose POCT (Nova Biomedical; Waltham, MA, USA).

The results of glucose values covered the range of hypoglycemia to hyperglycemia, with glucose concentrations ranging from 13 mg/dL to 389 mg/dL. The results obtained for the Nova StatStrip Glucose POCT were not affected by the level of hematocrit, by sample pH or by medication. In the two children’s hospitals where the study was performed, the NICU patient population demonstrated a wide range of hematocrit levels even beyond the levels of the ranges claimed as not showing interferences in the product labeling of other glucose meters.

The authors concluded that the accuracy of the NOVA StatStrip Glucose meter results was not influenced by the varying hematocrit levels in the NICU patients, which validates the performance of the hematocrit correction that is in-built into the design of StatStrip Glucose. The StatStrip meter showed good clinical accuracy and performance for measuring and monitoring glucose levels in NICU patients, with special respect to preterm infants, and therefore can act as a perfect alternative to a blood gas analyzer for measuring blood glucose in NICU patients. The study was published on September 1, 2013, in the journal Clinical Chemistry and Laboratory Medicine.

Related Links:
Medical University of Vienna
Radiometer Medical ApS
Nova Biomedical


Gold Member
Pharmacogenetics Panel
VeriDose Core Panel v2.0
Verification Panels for Assay Development & QC
Seroconversion Panels
New
Fecal DNA Extraction Kit
QIAamp PowerFecal Pro DNA Kit
New
Piezoelectric Micropump
Disc Pump
Read the full article by registering today, it's FREE! It's Free!
Register now for FREE to LabMedica.com and get complete access to news and events that shape the world of Clinical Laboratory Medicine.
  • Free digital version edition of LabMedica International sent by email on regular basis
  • Free print version of LabMedica International magazine (available only outside USA and Canada).
  • Free and unlimited access to back issues of LabMedica International in digital format
  • Free LabMedica International Newsletter sent every week containing the latest news
  • Free breaking news sent via email
  • Free access to Events Calendar
  • Free access to LinkXpress new product services
  • REGISTRATION IS FREE AND EASY!
Click here to Register








Channels

Molecular Diagnostics

view channel
Image: The study investigated D-dimer testing in patients who are at higher risk of pulmonary embolism (Photo courtesy of Adobe Stock)

D-Dimer Testing Can Identify Patients at Higher Risk of Pulmonary Embolism

Pulmonary embolism (PE) is a commonly suspected condition in emergency departments (EDs) and can be life-threatening if not diagnosed correctly. Achieving an accurate diagnosis is vital for providing effective... Read more

Immunology

view channel
Image: The findings were based on patients from the ADAURA clinical trial of the targeted therapy osimertinib for patients with NSCLC with EGFR-activated mutations (Photo courtesy of YSM Multimedia Team)

Post-Treatment Blood Test Could Inform Future Cancer Therapy Decisions

In the ongoing advancement of personalized medicine, a new study has provided evidence supporting the use of a tool that detects cancer-derived molecules in the blood of lung cancer patients years after... Read more

Microbiology

view channel
Image: Schematic representation illustrating the key findings of the study (Photo courtesy of UNIST)

Breakthrough Diagnostic Technology Identifies Bacterial Infections with Almost 100% Accuracy within Three Hours

Rapid and precise identification of pathogenic microbes in patient samples is essential for the effective treatment of acute infectious diseases, such as sepsis. The fluorescence in situ hybridization... Read more
Copyright © 2000-2025 Globetech Media. All rights reserved.