Hemocytometers Evaluated for Peritoneal Fluid Cell Counts

By LabMedica International staff writers
Posted on 21 Jan 2013
Print article
Image: The XE-5000 analyzer (Photo courtesy of Sysmex).
Image: The XE-5000 analyzer (Photo courtesy of Sysmex).
The diagnostic performance of different automated hemocytometers for white blood cell (WBC) enumeration of peritoneal fluids was compared with manual microscopy analysis.

The term ascites is conventionally used for designating the accumulation of fluid in the peritoneal cavity and the presence of elevate numbers of WBCs and polymorphonucleated leukocytes (PMNs) is pivotal to diagnose spontaneous bacterial peritonitis (SBP).

Scientists at the Academic Hospital of Parma (Italy) analyzed 100 peritoneal fluids with manual microscopy and five automated hemocytometers. Manual microscopic analysis was carried out using Nageotte and Fuchs-Rosenthal chambers after staining with Turk's and May-Grünwald–Giemsa reagents.

The five automated counters were the XE-2100 and XE-5000 (Sysmex; Kobe, Japan); the Advia 2120 (Siemens; Erlangen Germany); the BC-6800 (Mindray; Mahwah, NJ, USA); and the Abbott Sapphire (Abbott Park, IL, USA). The XE-2100 is a flow cytometer, using forward-scattered and side-scattered light, whereas the WBC differential entails a specific nucleic acid dye to measure the cells by side-fluorescent light and side-scattered light. The analysis of WBC as performed on the XE-5000 is with a flow cytometry technique by means of a semiconductor laser and fluorescent measurement.

For manual microscopy the mean values in the 100 peritoneal fluids was 873 WBC/mm3, compared to 943 WBC/mm3 for the XE-2100 and 741 WBC/mm3 for XE-5000. For the Advia 2120, the mean was 741 WBC/mm3, while for the BC-6800 it was 938 WBC/mm3 and 782 WBC/mm3 for the Sapphire. There were no significant differences between the different counts and therefore correlations were highly significant. The agreement between manual analysis and flow cytometry at the diagnostic threshold for septic peritonitis with a count equal to or greater than 1000 WBC/mm3, was 99% for XE-2100, XE-5000, Advia 2120, BC-6800 and 96% for Sapphire.

The authors concluded that used for routine analysis of body fluids the five hemocytometers tested display acceptable performance for routine screening of peritoneal fluid. Higher correlations with manual microscopy were found for the two Sysmex analyzers although those of the Advia 2120, BC-6800 and Sapphire were still excellent. The study was published in the January 2013 issue of the journal Clinical Biochemistry.

Related Links:

Academic Hospital of Parma
Sysmex
Siemens


Gold Member
Turnkey Packaging Solution
HLX
Antipsychotic TDM AssaysSaladax Antipsychotic Assays
New
Silver Member
Benchtop Image Acquisition Device
Microwell Imager
New
Free Human Prostate-Specific Antigen CLIA
LIAISON fPSA

Print article

Channels

Hematology

view channel
Image: The smartphone technology measures blood hemoglobin levels from a digital photo of the inner eyelid (Photo courtesy of Purdue University)

First-Of-Its-Kind Smartphone Technology Noninvasively Measures Blood Hemoglobin Levels at POC

Blood hemoglobin tests are among the most frequently conducted blood tests, as hemoglobin levels can provide vital insights into various health conditions. However, traditional tests are often underutilized... Read more

Immunology

view channel
Image: Under a microscope, DNA repair is visible as bright green spots (“foci”) in the blue-stained cell DNA. Orange highlights actively growing cancer cells (Photo courtesy of WEHI)

Simple Blood Test Could Detect Drug Resistance in Ovarian Cancer Patients

Every year, hundreds of thousands of women across the world are diagnosed with ovarian and breast cancer. PARP inhibitors (PARPi) therapy has been a major advancement in treating these cancers, particularly... Read more

Microbiology

view channel
Image: HNL Dimer can be a novel and potentially useful clinical tool in antibiotic stewardship in sepsis (Photo courtesy of Shutterstock)

Unique Blood Biomarker Shown to Effectively Monitor Sepsis Treatment

Sepsis remains a growing problem across the world, linked to high rates of mortality and morbidity. Timely and accurate diagnosis, along with effective supportive therapy, is essential in reducing sepsis-related... Read more
Copyright © 2000-2024 Globetech Media. All rights reserved.