Arterial Gas Analyzer and Autoanalyzer Compared for Electrolytes

By LabMedica International staff writers
Posted on 28 Mar 2013
Print article
Electrolyte values are measured in most critically ill intensive care unit (ICU) patients using both an arterial blood gas analyzer (ABG) and a central laboratory autoanalyzer (AA).

Data on sodium and potassium ion concentrations in ICU patients have been examined to see whether the electrolyte levels assessed by the ABG and the AA are equivalent and reliable.

Scientists at the Sevket Yilmaz Education and Research Hospital (Bursa, Turkey) retrospectively studied patients hospitalized in the ICU between July and August 2011. They identified 84 instances, of 1,105 patient blood gas samples analyzed, in which arterial and venous samples were collected simultaneously and sodium (Na+) and (K+) were measured using two methods.

In the ICU, arterial blood samples were collected in heparinized blood-gas syringes and analyzed using the benchtop pHOx Stat Profile Plus L blood–gas analyzer (Nova Biomedical; Waltham MA, USA) which employs direct ion-specific electrode (ISE) technology. Another sample was drawn, at the same time, from the same arterial sampling point, using a vacuum technique featuring clot-activating tubes. These samples were sent, pneumatically sealed, to the central laboratory, where serum Na+ and K+ concentrations were analyzed via indirect ISE on a Modular ISE900 auto-analyzer (Roche Diagnostics; Mannheim, Germany).

The mean sodium concentration was 138.mmol/L (standard deviation, SD ±10.2 mmol/L) using the ABG and 143.0 mmol/L (SD ±10.5 mmol/L) using the AA, which was statistically different. The mean potassium level was 3.5mmol/L (SD ± 0.9 mmol/L) using the ABG and 3.7 mmol/L (SD ± 1.0 mmol/L) using the AA which was also significantly different. The extent of interanalyzer agreement was unacceptable for both K+ and Na+. The between-day imprecision of both instruments (AA and ABG) was small and lacked clinical significance when compared with analytical performance indicators based on biological variation.

The authors concluded that their study illustrates the importance of determining the concordance, for each individual hospital, of electrolyte values obtained by ABG and those obtained in the central laboratory. As instrument type and calibration methods may differ among hospitals, it is important that each center conducts an in-house study as in their study the ABG and AA do not yield equivalent Na+ and K+ data. The study was published on August 3, 2012, in the BioMed Central journal BMC Anesthesiology.

Related Links:

Sevket Yilmaz Education and Research Hospital
Nova Biomedical
Roche



Gold Member
Blood Gas Analyzer
GEM Premier 7000 with iQM3
New
Gold Member
LEISHMANIA Test
LEISHMANIA ELISA
New
Blood Gas Panel plus Electrolytes
i-STAT EG6+ Cartridge
New
Ureaplasma Urealyticum Test
Duplicα RealTime Ureaplasma Urealyticum Kit

Print article

Channels

Molecular Diagnostics

view channel
Image: A coronal MRI section shows a high-intensity focused ultrasound lesion in the left thalamus of the brain (Photo courtesy of UT Southwestern Medical Center)

Newly Identified Stroke Biomarkers Pave Way for Blood Tests to Quickly Diagnose Brain Injuries

Each year, nearly 800,000 individuals in the U.S. experience a stroke, which occurs when blood flow to specific areas of the brain is insufficient, causing brain cells to die due to a lack of oxygen.... Read more

Immunology

view channel
Image: The discovery of biomarkers could improve endometrial cancer treatment (Photo courtesy of Mount Sinai)

Simple Blood Test Could Help Choose Better Treatments for Patients with Recurrent Endometrial Cancer

Endometrial cancer, which develops in the lining of the uterus, is the most prevalent gynecologic cancer in the United States, affecting over 66,000 women annually. Projections indicate that in 2025, around... Read more

Microbiology

view channel
Image: The study identified a genetic signature in bacteria that, when present, indicates the likelihood of developing antibiotic resistance (Photo courtesy of Tulane University)

Unique Genetic Signature Predicts Drug Resistance in Bacteria

Antibiotic resistance represents a significant global health threat, responsible for over a million deaths each year. By 2050, the World Health Organization predicts that it could surpass cancer and heart... Read more

Pathology

view channel
Image: As tumor cells flow through these microfluidic chambers, they are subjected to increasing shear stress and sorted based on their adhesion strength (Photo courtesy of UC San Diego)

Microfluidic Device Assesses Stickiness of Tumor Cells to Predict Cancer Spread

Ductal carcinoma in situ (DCIS), a type of early-stage breast cancer, is often referred to as stage zero breast cancer. In many cases, it remains harmless and does not spread beyond the milk ducts where... Read more
Copyright © 2000-2025 Globetech Media. All rights reserved.