Blood Glucose Monitoring System Designed for Accuracy

By LabMedica International staff writers
Posted on 29 Jul 2013
Recently more attention has been drawn to the accuracy requirements for blood glucose monitoring systems (BGMS) for home use as well as those for hospital use.

Tighter accuracy criteria are being introduced internationally for BGMS and therefore newer systems need to be evaluated for point-of-care testing (POCT) of blood glucose, a valuable biomarker for the management of patients with diabetes and other conditions. More...


Collaborating scientists from three different institutions and the commercial manufacturer evaluated a BGMS in a multicenter study for precision, interference, and system accuracy using 503 blood samples from capillary, venous, and arterial sources. Included in the study was interference testing, both hematocrit and other substances, and accuracy evaluations. The institutes were the Rainier Clinical Research Center (Renton, WA, USA), the Duke University Medical Center (Durham, NC, USA), and MassResearch LLC (Waltham, MA, USA).

The FreeStyle Precision Pro Blood Glucose and β-Ketone Monitoring System (Abbott Diabetes Care; Alameda, CA, USA) is designed to simplify POCT for healthcare professionals, providing features that enhance the reliability of the testing process and support compliance with POCT policies. The FreeStyle Precision Pro blood glucose-test strip is individually sealed, and each test strip contains three electrodes, working, reference, and fill trigger. The circuit between the fill trigger and the reference electrodes must be detected by the meter before the test will start. The minimum sample volume required is 0.6 μL and the test time is five seconds.

The FreeStyle Precision Pro system demonstrated acceptable precision, no interference across a hematocrit range of 15% to 65%, and, except for xylose, no interference from 24 of 25 potentially interfering substances. From the individual results of various blood sample types, 97.3% to 98.9% agreed within ±12 mg/dL of the laboratory analyzer values at glucose concentrations less than 100 mg/dL and within ±12.5% of the laboratory analyzer values at glucose concentrations equal to or greater than 100 mg/dL.

The authors concluded that the FreeStyle Precision Pro system makes more measurements and uses more signals from the glucose electrochemical reaction to increase accuracy; it also performs more integrity checks to detect potential errors during each glucose test. The study was published on June 5, 2013, in the journal Clinica Chimica Acta.

Related Links:

Rainier Clinical Research Center
Duke University Medical Center
Abbott Diabetes Care



Gold Member
Immunochromatographic Assay
CRYPTO Cassette
Portable Electronic Pipette
Mini 96
HBV DNA Test
GENERIC HBV VIRAL LOAD VER 2.0
New
Gold Member
Hybrid Pipette
SWITCH
Read the full article by registering today, it's FREE! It's Free!
Register now for FREE to LabMedica.com and get access to news and events that shape the world of Clinical Laboratory Medicine.
  • Free digital version edition of LabMedica International sent by email on regular basis
  • Free print version of LabMedica International magazine (available only outside USA and Canada).
  • Free and unlimited access to back issues of LabMedica International in digital format
  • Free LabMedica International Newsletter sent every week containing the latest news
  • Free breaking news sent via email
  • Free access to Events Calendar
  • Free access to LinkXpress new product services
  • REGISTRATION IS FREE AND EASY!
Click here to Register








Channels

Hematology

view channel
Image: A schematic illustrating the coagulation cascade in vitro (Photo courtesy of Harris, N., 2024)

ADLM’s New Coagulation Testing Guidance to Improve Care for Patients on Blood Thinners

Direct oral anticoagulants (DOACs) are one of the most common types of blood thinners. Patients take them to prevent a host of complications that could arise from blood clotting, including stroke, deep... Read more

Microbiology

view channel
Image: EBP and EBP plus have received FDA 510(k) clearance and CE-IVDR Certification for use on the BD COR system (Photo courtesy of BD)

High-Throughput Enteric Panels Detect Multiple GI Bacterial Infections from Single Stool Swab Sample

Gastrointestinal (GI) infections are among the most common causes of illness worldwide, leading to over 1.7 million deaths annually and placing a heavy burden on healthcare systems. Conventional diagnostic... Read more
Copyright © 2000-2025 Globetech Media. All rights reserved.