Rapid Method Detects Urinary Tract Infections

By LabMedica International staff writers
Posted on 22 May 2012
The reliability of a rapid detection method for urinary tract infection (UTI) has been compared with the reference standard streaked agar plate method. More...


The novel CultureStat Rapid UTI Detection System (CSRUDS) consists of a vacuum-sealed glass ampoule containing trypticase soy broth dry media and triphenyl tetrazolium chloride, a chromophore indicator sensitive to dehydrogenase.

Scientists at Columbia University Medical Center, (New York, NY, USA) working with others, selected 980 clean catch urine samples into specimen cups during a 30-day period. Each specimen was streaked on separate 5% sheep blood agar plates and simultaneously a 7-mL aliquot of each fresh urine specimen was processed using the CSRUDS. A spectrophotometer linked to a computer and a bar code scanner was used to identify and test each urine sample.

Positive UTI conditions were accurately identified by both CSRUDS and agar streak plate methods. CSRUDS accurately identified UTI negative conditions with 99.3% reliability at 90 minutes. The negative predictive value of CSRUDS was 99.2% at 30 minutes. The bacterial plating results revealed that 413 (58.8%) of the CSRUDS negative group demonstrated no bacterial growth at all, and 289 (41.2%) of the CSRUDS negative group showed some degree of bacterial growth. Of the 289 samples, 82 (28.4%, including one of the six CSRUDS false-negative samples), showed bacterial growth in the 50,000 colony forming units (CFU) range and 200/289 (69.2%), including the other five false-negative CSRUDS samples, had CFU counts of 1,000-49,000 CFUs or mixed flora. Of the 289, 7 (2.4%) had growth greater than 100,000 colonies/mL.

The author concluded that he novel rapid detection system evaluated in this test program, the CSRUDS and MB80CSENC spectrophotometer (MicroBionetics, Eldersburg, MD, USA) provided UTI negative/positive same-day results within 90 minutes from the start of the test, with a highly reliable negative predictive value within 30 minutes for most patients. Because it is such a rapid test, contamination at urine collection is unlikely to affect the results of the assay. Specimen transport and delays in processing the urine sample are obviated because the test can be immediately performed in the urologist's office where the specimen was collected. The study was published in the April 2012 issue of the journal Urology.

Related Links:
Columbia University Medical Center
MicroBionetics


Platinum Member
Automated Coagulation Analyzer
Hemolumi H6
Gold Member
Quantitative POC Immunoassay Analyzer
EASY READER+
New
Slide Scanner System
NanoZoomer S540MD
Gold Member
Platelet Function Analyzer
PL-12
Read the full article by registering today, it's FREE! It's Free!
Register now for FREE to LabMedica.com and get access to news and events that shape the world of Clinical Laboratory Medicine.
  • Free digital version edition of LabMedica International sent by email on regular basis
  • Free print version of LabMedica International magazine (available only outside USA and Canada).
  • Free and unlimited access to back issues of LabMedica International in digital format
  • Free LabMedica International Newsletter sent every week containing the latest news
  • Free breaking news sent via email
  • Free access to Events Calendar
  • Free access to LinkXpress new product services
  • REGISTRATION IS FREE AND EASY!
Click here to Register








Channels

Clinical Chemistry

view channel
Image Credit: 123RF

Screening After Pneumococcal Disease May Reveal Undiagnosed Blood Cancer or Immune Disorders

Severe pneumococcal disease requiring hospitalization often presents as pneumonia, particularly in older adults and people with cancer or compromised immune function. Because M protein testing and antibody... Read more

Technology

view channel
Image: The laser-based photoacoustic spectroscopy setup consists of a Mid-IR laser equipped with three QCL modules covering wavelengths from 5.6 μm to 12.9 μm, two silver coated mirrors (SCM), a dichroic mirror (DM) with a transmittance of 90%, a thermal power sensor head (PM) to monitor the output laser power, a mechanical chopper (MC) for frequency modulation and a CEPAS-detector with a self-designed swab holder (SH). (Credit: Graunke, T., Scholz, T., Pieniak, M. et al. Scientific Reports (2026). https://doi.org/10.1038/s41598-026-68298-9)

Laser-Based Swab Analysis Shows Promise for Detecting Disease-Linked Odor Patterns

Disease-related changes in volatile organic compounds can alter body odor, producing measurable patterns in exhaled breath and bodily fluids. Current analytical methods can be complex, time-consuming,... Read more

Industry

view channel
Image

Collaboration Combines AI Cognitive Assessment and RNA Blood Testing for Earlier Alzheimer’s Detection

Alzheimer’s disease is often identified only after substantial neurodegeneration, partly because current diagnostic pathways are fragmented and difficult to scale. As treatment shifts toward earlier intervention,... Read more
Copyright © 2000-2026 Globetech Media. All rights reserved.