Rapid Diagnostic Test Evaluated for Human Malaria Species

By LabMedica International staff writers
Posted on 04 Jul 2012
Print article
A rapid diagnostic test (RDT) for malaria parasites has been evaluated and compared with other methods in a reference setting.

The three-band rapid diagnostic test detects Plasmodium falciparum-specific parasite lactate dehydrogenase (Pf-pLDH) and pan Plasmodium-specific pLDH (pan-pLDH).

Scientists at Maastricht University (Netherlands) carried out a retrospective and a prospective study to assess with 498 stored and 77 fresh blood samples obtained from international travelers suspected of malaria. Both panels comprised all four Plasmodium species; the retrospective panel comprised also samples negative for malaria. The retrospective panel had been obtained between February 1996 and May 2011. The reference method was microscopy corrected by polymerase chain reaction (PCR). The prospective panel was run side-to-side with two other RDTs.

In the retrospective evaluation, overall sensitivity for the 247 P. falciparum samples was 94.7%, reaching 98.7% for parasite densities greater than 1,000/μL. Most false negative results occurred among samples with pure gametocytaemia and at parasite densities less than or equal to 100/μL. None of the 96 negative samples showed visible test lines. Sensitivities for the 70 P. vivax, 69 P. ovale and 16 P. malariae were 74.3%, 31.9% and 25.0% respectively. Overall, Pf-pLDH test lines showed higher line intensities compared to the pan-pLDH lines, with 67.9% and 23.0% medium and strong line intensities for P. falciparum.

In the prospective panel, the CareStart pLDH RDT (AccessBio Inc.; Monmouth, NJ, USA) showed higher sensitivities for P. falciparum compared to OptiMAL (Pf-pLDH/pan-pLDH, BioRad; Marnes-la-Coquette, France) and lower sensitivities for P. falciparum as compare to SDFK60 (histidine-rich protein-2 (HRP-2)/pan-pLDH; Standard Diagnostics; Hagal-Dong, Korea). Wrong species identification occurred in 10 (2.5%) samples and was mainly due to P. vivax samples reacting with the Pf-pLDH test line.

The authors concluded that CareStart pLDH may be an alternative to other well-described HRP-2 three band RDTs in the diagnostic setting in non-endemic areas. Despite being better than the other two RDTs that were run side by side, it should be noted that its performance for the detection of P. ovale and P. malariae is mediocre. The study was published on June 18, 2012, in the Malaria Journal.

Related Links:

Maastricht University
AccessBio Inc.
BioRad


Gold Member
C-Reactive Protein Reagent
CRP Ultra Wide Range Reagent Kit
Automated Blood Typing System
IH-500 NEXT
New
Lab Sample Rotator
H5600 Revolver
New
Gold Member
Pharmacogenetics Panel
VeriDose Core Panel v2.0

Print article

Channels

Clinical Chemistry

view channel
Image: The new saliva-based test for heart failure measures two biomarkers in about 15 minutes (Photo courtesy of Trey Pittman)

POC Saliva Testing Device Predicts Heart Failure in 15 Minutes

Heart failure is a serious condition where the heart muscle is unable to pump sufficient oxygen-rich blood throughout the body. It ranks as a major cause of death globally and is particularly fatal for... Read more

Molecular Diagnostics

view channel
Image: This medium is used to grow malaria parasites (Photo courtesy of Kyle Dykes/UC San Diego Health Sciences)

New Approach to Help Predict Drug Resistance in Malaria and Infectious Diseases

Malaria, a disease transmitted by mosquitoes that affects millions worldwide, remains a significant public health concern, especially in tropical and subtropical regions. Despite significant efforts to... Read more

Hematology

view channel
Image: The smartphone technology measures blood hemoglobin levels from a digital photo of the inner eyelid (Photo courtesy of Purdue University)

First-Of-Its-Kind Smartphone Technology Noninvasively Measures Blood Hemoglobin Levels at POC

Blood hemoglobin tests are among the most frequently conducted blood tests, as hemoglobin levels can provide vital insights into various health conditions. However, traditional tests are often underutilized... Read more

Immunology

view channel
Image: Under a microscope, DNA repair is visible as bright green spots (“foci”) in the blue-stained cell DNA. Orange highlights actively growing cancer cells (Photo courtesy of WEHI)

Simple Blood Test Could Detect Drug Resistance in Ovarian Cancer Patients

Every year, hundreds of thousands of women across the world are diagnosed with ovarian and breast cancer. PARP inhibitors (PARPi) therapy has been a major advancement in treating these cancers, particularly... Read more
Copyright © 2000-2024 Globetech Media. All rights reserved.