POC Testing Compared to Automated Blood Glucose Analysis

By LabMedica International staff writers
Posted on 13 Sep 2016
Print article
Image: The OneTouch UltraVue POC blood glucose monitor (Photo courtesy of Johnson & Johnson).
Image: The OneTouch UltraVue POC blood glucose monitor (Photo courtesy of Johnson & Johnson).
Diabetes mellitus is a metabolic disease that is characterized by hyperglycemia and blood glucose (BG) levels are helpful for the diagnosis and treatment of diabetes and an important part of the management of diabetes.

Point-of-care testing (POCT) is generally used by patients themselves or medical personnel to monitor BG. As it is simple to use and supplies rapid results, it might accelerate clinical decision-making and improve patients' prognosis.

Scientists at the Guangxi Medical University (Nanning, China) carried out a retrospective study comprising 162 volunteers (97 males and 65 females; median age, 47 years; range, 14 to 67 years). Blood samples were collected with anticoagulation tubes for which hematocrit values ranged between 34.5% and 54% and the range of glucose concentrations were 1.60–21.30 mmol/L. Each sample was divided into two parts; one was separated by centrifugation and measured by the central laboratory system, and the other was measured by the POCT system in whole blood.

The team tested three types of BG meter: They were the ACCU-CHEK Performa and the ACCU-CHEK Active (Roche Diagnostics, Basel, Switzerland), and the OneTouch UltraVue (Johnson & Johnson Medical Ltd, Shanghai, China). The clinical laboratory measurement system used the Glucose GOD-PAP method with a Hitachi 7600-120 analyzer (Hitachi Corporation, Tokyo, Japan) as a reference system.

The investigators showed that BG concentrations measured by venous whole blood was comparable with the reference method. Although the POCT can provide rapid measurement results, the type of specimen, interference factors, environmental conditions, and hematocrit may affect the results. The measurement accuracy of the POCT systems, being 94.9%, 92.6%, and 96.7% for ACCU-CHEK Performa, ACCU-CHEK Active, and OneTouch UltraVue, respectively, did not all meet the requirements. The same result was found for BG concentrations equal to or greater than 5.55 mmol/L (100 mg/dL). With BG concentrations of less than 5.55 mmol/l (100 mg/dL), the accuracy of all POCT systems meet the criteria.

The authors concluded that their study indicated that the POCT system could only be used as a screening test for continuous, qualitative, or semi-quantitative detection as a complement emergency measure to the central laboratory, instead of being used as a standard result in clinical diagnosis. If the POCT system is used in testing too low or too high BG concentrations in clinical diagnosis, it may mislead the diagnosis and treatment. The study was published on August 25, 2016, in the Journal of Clinical Laboratory Analysis.

Related Links:
Guangxi Medical University
Roche Diagnostics
Johnson & Johnson Medical
Hitachi Corporation
Gold Member
Antipsychotic TDM Assays
Saladax Antipsychotic Assays
Verification Panels for Assay Development & QC
Seroconversion Panels
New
Aspergillus Test
REALQUALITY Aspergillus
New
Immunofluorescence Analyzer
MPQuanti

Print article

Channels

Immunology

view channel
Image: The cancer stem cell test can accurately choose more effective treatments (Photo courtesy of University of Cincinnati)

Stem Cell Test Predicts Treatment Outcome for Patients with Platinum-Resistant Ovarian Cancer

Epithelial ovarian cancer frequently responds to chemotherapy initially, but eventually, the tumor develops resistance to the therapy, leading to regrowth. This resistance is partially due to the activation... Read more

Microbiology

view channel
Image: The lab-in-tube assay could improve TB diagnoses in rural or resource-limited areas (Photo courtesy of Kenny Lass/Tulane University)

Handheld Device Delivers Low-Cost TB Results in Less Than One Hour

Tuberculosis (TB) remains the deadliest infectious disease globally, affecting an estimated 10 million people annually. In 2021, about 4.2 million TB cases went undiagnosed or unreported, mainly due to... Read more

Technology

view channel
Image: The HIV-1 self-testing chip will be capable of selectively detecting HIV in whole blood samples (Photo courtesy of Shutterstock)

Disposable Microchip Technology Could Selectively Detect HIV in Whole Blood Samples

As of the end of 2023, approximately 40 million people globally were living with HIV, and around 630,000 individuals died from AIDS-related illnesses that same year. Despite a substantial decline in deaths... Read more

Industry

view channel
Image: The collaboration aims to leverage Oxford Nanopore\'s sequencing platform and Cepheid\'s GeneXpert system to advance the field of sequencing for infectious diseases (Photo courtesy of Cepheid)

Cepheid and Oxford Nanopore Technologies Partner on Advancing Automated Sequencing-Based Solutions

Cepheid (Sunnyvale, CA, USA), a leading molecular diagnostics company, and Oxford Nanopore Technologies (Oxford, UK), the company behind a new generation of sequencing-based molecular analysis technologies,... Read more
Copyright © 2000-2025 Globetech Media. All rights reserved.