Performance of Immunosuppressant Drug Assay Kits Analyzed

By LabMedica International staff writers
Posted on 08 Aug 2017
Print article
Image: The cobas e 411immunoassay analyzer (Photo courtesy of Roche Diagnostics).
Image: The cobas e 411immunoassay analyzer (Photo courtesy of Roche Diagnostics).
Cyclosporine (CsA) and tacrolimus (TAC) are immunosuppressant drugs that are often used to treat autoimmune diseases and as transplantation therapy and therefore, their concentrations need to be monitored carefully.

The main analytical methods for CsA and TAC in whole blood have been liquid chromatography combined with mass spectrometric detection (LC-MS/MS), affinity chrome-mediated immunoassay (ACMIA), and chemiluminescence immunoassays (CLIA).

Scientists at the Osaka University Hospital (Osaka, Japan) used 200 residual EDTA whole blood samples from patients who had received an organ transplant and were under CsA or TAC therapy. CsA concentrations were measured using an affinity chrome-mediated immunoassay (ACMIA) and an electrochemiluminescence immunoassay (ECLIA). TAC concentrations were measured using a chemiluminescence immunoassay (CLIA) and ECLIA.

The investigators used the Elecsys Cyclosporine assay kit and the Elecsys Tacrolimus assay kit where the concentrations were measured on a cobas e411 analyzer. The ACMIA for measurement of CsA concentration was performed with a Dimension Xpand analyzer and CLIA for measurement of TAC concentration with an ARCHITECT i2000SR. It was necessary for accurate measurements of CsA and TAC concentrations to sufficiently mix samples and pretreatment reagent. The hematocrit was assayed using the fully automated hematological analyzer Sysmex XN-1000.

The scientists investigated assay precision, linearity, lower limit of quantitation (LOQ), stability of calibration, influence of interference substances and the hematocrit, correlation of ACMIA with ECLIA, and correlation of CLIA with ECLIA. They found that the limits of quantitation (LOQ) were defined as the concentration at which the coefficient of variation (CV) was approximately 10%. Each lower LOQ obtained was 16 ng/mL (CsA), and 0.95 ng/mL (TAC). CsA and TAC calibrations were stable for at least 21 days. Neither the presence of conjugated bilirubin, unconjugated bilirubin, chyle, and rheumatoid factor nor the hematocrit affected these assays.

The authors concluded that the analytical performances of the Elecsys Cyclosporine and Elecsys Tacrolimus assays were acceptable. Furthermore, CsA and TAC concentrations may be simultaneously measured using a single pretreatment, which is of benefit if patients have to undertake conversion between these two drugs. Additionally, it benefits the workflow in the clinical laboratory. The study was published in the August 2017 issue of the journal Practical Laboratory Medicine.

Gold Member
Chagas Disease Test
CHAGAS Cassette
Verification Panels for Assay Development & QC
Seroconversion Panels
New
Coagulation Analyzer
CS-2400
New
Respiratory QC Panel
Assayed Respiratory Control Panel

Print article

Channels

Molecular Diagnostics

view channel
Image: The experimental blood test accurately indicates severity and predicts potential recovery from spinal cord injury (Photo courtesy of 123RF)

Blood Test Identifies Multiple Biomarkers for Rapid Diagnosis of Spinal Cord Injury

The National Institutes of Health estimates that 18,000 individuals in the United States sustain spinal cord injuries (SCIs) annually, resulting in a staggering financial burden of over USD 9.... Read more

Microbiology

view channel
Image: Schematic representation illustrating the key findings of the study (Photo courtesy of UNIST)

Breakthrough Diagnostic Technology Identifies Bacterial Infections with Almost 100% Accuracy within Three Hours

Rapid and precise identification of pathogenic microbes in patient samples is essential for the effective treatment of acute infectious diseases, such as sepsis. The fluorescence in situ hybridization... Read more

Industry

view channel
Image: Tumor-associated macrophages visualized using the Multiomic LS Assay (Photo courtesy of ACD)

Leica Biosystems and Bio-Techne Expand Spatial Multiomic Collaboration

Bio-Techne Corporation (Minneapolis, MN, USA) has expanded the longstanding partnership between its spatial biology brand, Advanced Cell Diagnostics (ACD, Newark, CA, USA), and Leica Biosystems (Nussloch,... Read more
Copyright © 2000-2025 Globetech Media. All rights reserved.